Blister
packaging is transparent in appearance, clear in display, convenient for usage
and light in weight; thus, it has been widely applied in solid pharmaceutical
packaging as an effective protective means for contents. Blister packaging need
to have very excellent oxygen barrier property so as to prevent deterioration
during validity period. Thus, it has
been clearly stipulated in the relevant standards for the oxygen transmission
property, which is a key index for manufacturers. Hereinafter,
this article is to briefly introduce the oxygen transmission
property test of blister packaging with Labthink OX23/230
Oxygen Transmission Rate Test System.
Blister
packaging refers to the packaging means that seals the content in between a
preformed transparent plastic blister and a backing of paperboard, plastic
film, plastic sheet, aluminum foil, or compounds, etc.
Oxygen transmission property of blister packaging is referred to its
barrier property against oxygen, which is an important factor that could
influence shelf life of blister packed pharmaceuticals thus provides a critical
reference to the shelf life analysis. Through oxygen transmission property test, oxidization and deterioration
problems caused by oxygen sensitiveness could be evaluated.
Coulometric method
is adopted by Labthink OX2/230 Oxygen Transmission Rate TestSystem. The
working principle of Labthink OX2/230
Oxygen Transmission Rate Test System is as follows: place the
preconditioned blister package specimen into the testing chamber,
then,
high pure nitrogen flows in the package; air or pure oxygen flows outside the
package. Owing to the condensation difference of oxygen, oxygen molecules
permeate through the specimen and enter into the high pure nitrogen side, which
is then carried to the sensor by the flowing nitrogen. Then the oxygen
transmission rate could be obtained by analyzing the oxygen condensation
detected by the sensor and the oxygen transmission property of the specimen
could be evaluated.
Oxygen Transmission Rate Test System Principle |
Labthink OX2/230 Oxygen Transmission Rate Test System is
fully automatic, controlled by computer and it has 3 testing chambers
integrated in one block (Labthink patent). Clamp the specimen, click Test
button, and the test will be started automatically with auto control of
temperature, humidity, flow rate and other indexes, then the test system will
automatically judge the test status and provide with test report when testing
is competed.
Ox2/230 Oxygen Transmission Rate Test System |
Labthink OX2/230 could connect 9 satellite modules connect
9 satellite modules, which could be the Oxygen Transmission Rate Test System
and Water Vapor Transmission Rate Test System, thus it could process 30
specimens simultaneously. It maintains a high level in automation, test
efficiency, detection resolution and lower limit of test range, which makes
Labthink OX2/230 rank in a leading position in this industry worldwide.
Besides oxygen transmission
property, more indexes such as water vapor transmission property and sealing performance, etc., need to be tested for blister packaging
to guarantee quality of pharmaceuticals within the shelf life. For more
information of other related testing instruments, please visit http://www.labthinkinternational.com.cn.